SD Card inside Analyzer is Corrupted Please contact your POCT for replacement
The firmware has corrupted the SD card, or a power error has occurred.
- Restart Handheld and attempt new test with new Cartridge.
- If error still occurs, Analyzer be returned to NanoDx™ immediately for further inspection.
Related Articles
No SD card present in Analyzer. Please contact your POCT
SD card was not detected in the Analyzer. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again.) Attempt new test with new Cartridge If error still occurs, Analyzer should be returned to NanoDx™ for further ...
A thread-related error has occurred in the Analyzer. Please contact your POCT:
This is an occasional firmware error. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again). Attempt new test with new Cartridge. If error still occurs, Analyzer should be returned to NanoDx™.
A PWG calibration error has occurred in the Analyzer. Please contact your POCT:
This typically indicates a sensor error. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again). Attempt new test with new Cartridge. If error still occurs, Analyzer should be returned to NanoDx™.
An error has occurred with the Cartridge Sensors in the Analyzer. Please contact your POCT:
This error typically cannot be recovered from. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again). Attempt new test with new Cartridge. Iff error still occurs, Analyzer should be returned to NanoDx™.
An error has occurred in the Analyzer. Please contact your POCT:
Typically indicates a potential issue with the firmware. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again). Attempt new test with new Cartridge. If error still occurs, Analyzer should be returned to NanoDx™.