It is recommended that the NanoDx™ System functionality is tested periodically before use, in accordance with your institution's quality policies and procedures.
It is recommended that the NanoDx™ System functionality is tested periodically before use, in accordance with your institution's quality policies and procedures.
After you press the "START TEST" button on the Handheld, you must wait over 60 seconds to allow the Analyzer to prepare for the test. When prompted, insert the ETC into the Analyzer. Be sure not to lift the Handheld until the test is completed.
Specimens must be run within 10 minutes of collection to ensure accurate readings by the Analyzer. Using a specimen collected more than 10 minutes prior to insertion in the Analyzer may not generate accurate test results.
Yes, the Handheld must be placed into the cradle of the Analyzer in order to run either QC or Patient tests. The Handheld will prompt when to place the Handheld into the Analyzer's cradle. Note that when the Handheld prompts for User ID, Patient ID, ...
If any QC test is a FAIL, then re-run the QC test for that specific control. If the second test is a PASS, then procedd to the next step of the QC test procedure. If this re-test was the last step in the process, and all other tests have also been ...