How do I get the Cartridge out of the Analyzer?
- After the test is complete, the Cartridge will automatically be ejected from the Analyzer.
- Do not attempt to place your fingers into the Cartridge slot of the Analyzer in order to manually eject the Cartridge.
Related Articles
What if the Cartridge gets stuck in the Analyzer?
Power cycling (turning the Analyzer off, waiting 5-10s, then turning it on again) will eject the Cartridge once the Analyzer reboots. Upon turning the Analyzer on again, it will take about 30-60s for the Analyzer to boot and the Cartridge to eject.
An error has occurred with the Cartridge Sensors in the Analyzer. Please contact your POCT:
This error typically cannot be recovered from. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again). Attempt new test with new Cartridge. Iff error still occurs, Analyzer should be returned to NanoDx™.
Do I need to put the Cartridge in the Handheld when transporting to the Analyzer?
No, the Cartridge does not need to be placed in the Handheld when being transported to the Analyzer. However, it is recommended for ease of transport.
What happens if I use the wrong Cartridge for the test?
The Analyzer will eject the Cartridge. You should re-run the test using the correct Cartridge that corresponds with the test being run.
No SD card present in Analyzer. Please contact your POCT
SD card was not detected in the Analyzer. Power cycle the Analyzer (Power off the Analyzer, wait 10 seconds, and power it on again.) Attempt new test with new Cartridge If error still occurs, Analyzer should be returned to NanoDx™ for further ...