Will water or other cleaning products damage the system?
- Users shouldn't put anything into the opening of the Analyzer port.
- It's unavoidable getting some dribbles of cleaning fluid into the front opening of the Analyzer during the wipe down cleaning we specify; this will not pose a problem with performance nor pose safety hazard.
Related Articles
Will NanoDx™ provide liquid controls?
NanoDx™ will not provide liquid controls.
How do I clean the system?
The Analyzer, Handheld, and ETC must be cleaned periodically, based on your institution's protocols. These components can be cleaned using an alcohol wipe, containing at least 70% isopropyl alcohol (IPA). Avoid directly spraying the unit with IPA or ...
How do I configure the system?
If the Handheld is docked on the Analyzer, lift it off. If the Handheld is not already on, press and hold the button on the right-hand side to switch it on. On the "Start" screen, press "SETTINGS". You will the be prompted to enter an administrator ...
At what temperature should Cartridges be stored?/ How do I store system components?
Store cartridges between 2 °C to 8 °C (36 °F to 46 °F). System components should all be stored in a clean and dry space at room temperature (18 °C to 25 °C/64 °F to 46 °F).
What if you fail QC testing two times in a row?
Two failed Liquid QC on the same control (positive/negative) locks the Handheld from performing patient tests. Four failed Electronic QC tests (two in a row per ETC) will lock the system from performing patient tests.