Will water or other cleaning products damage the system?

Will water or other cleaning products damage the system?

  1. Users shouldn't put anything into the opening of the Analyzer port.
  2. It's unavoidable getting some dribbles of cleaning fluid into the front opening of the Analyzer during the wipe down cleaning we specify; this will not pose a problem with performance nor pose safety hazard. 

    • Related Articles

    • Will NanoDx™ provide liquid controls?

      NanoDx™ will not provide liquid controls. 
    • How do I clean the system?

      The Analyzer, Handheld, and ETC must be cleaned periodically, based on your institution's protocols. These components can be cleaned using an alcohol wipe, containing at least 70% isopropyl alcohol (IPA). Avoid directly spraying the unit with IPA or ...
    • How do I configure the system?

      If the Handheld is docked on the Analyzer, lift it off. If the Handheld is not already on, press and hold the button on the right-hand side to switch it on. On the "Start" screen, press "SETTINGS". You will the be prompted to enter an administrator ...
    • At what temperature should Cartridges be stored?/ How do I store system components?

      Store cartridges between 2 °C to 8 °C (36 °F to 46 °F). System components should all be stored in a clean and dry space at room temperature (18 °C to 25 °C/64 °F to 46 °F).
    • What if you fail QC testing two times in a row?

      Two failed Liquid QC on the same control (positive/negative) locks the Handheld from performing patient tests.  Four failed Electronic QC tests (two in a row per ETC) will lock the system from performing patient tests.