First, ensure that you chose the positive control first. If you did choose the positive control first, the ensure that you dispensed approximately 75 microliters of the positive control liquid into the Cartridge using the provided Specimen Transfer ...
Users shouldn't put anything into the opening of the Analyzer port. It's unavoidable getting some dribbles of cleaning fluid into the front opening of the Analyzer during the wipe down cleaning we specify; this will not pose a problem with ...
Two failed Liquid QC on the same control (positive/negative) locks the Handheld from performing patient tests. Four failed Electronic QC tests (two in a row per ETC) will lock the system from performing patient tests.
Press "RESULTS" on the Main Menu and scroll through the results. All QC tests are marked with a magnifying glass and will have the type (liquid or electronic) listed as well. Green row = pass, red row = invalid result or failure.