Do I need to clean the system in between patient tests?
Yes, you must clean both the Handheld and the Analyzer between patient tests and after positive and negative external control testing.
Related Articles
How do I clean the system?
The Analyzer, Handheld, and ETC must be cleaned periodically, based on your institution's protocols. These components can be cleaned using an alcohol wipe, containing at least 70% isopropyl alcohol (IPA). Avoid directly spraying the unit with IPA or ...
Does the Handheld need to be on the Analyzer when I run the test?
Yes, the Handheld must be placed into the cradle of the Analyzer in order to run either QC or Patient tests. The Handheld will prompt when to place the Handheld into the Analyzer's cradle. Note that when the Handheld prompts for User ID, Patient ID, ...
Where do I get the patient ID?
The patient ID will be consistent with their hospital identification.
At what temperature should Cartridges be stored?/ How do I store system components?
Store cartridges between 2 °C to 8 °C (36 °F to 46 °F). System components should all be stored in a clean and dry space at room temperature (18 °C to 25 °C/64 °F to 46 °F).
Where are patient test results dispayed?
Patient results are displayed on the Handheld after the test is completed. You can find patient results on the "Results" screen on the home screen of the Handheld. Patient test results are also available in the NanoDx™ Customer Portal.