Press the print icon button on any QC or Patient Test Result screen. If a printer has been configured the printer should start to print the test results within 60s.
Patient results are displayed on the Handheld after the test is completed. You can find patient results on the "Results" screen on the home screen of the Handheld. Patient test results are also available in the NanoDx™ Customer Portal.
Please refer to the operator's manual and re-run the test for that specific control. Possible sources for a shift in results include: Analyzer is near a source of inconsistent heat/cold (fans, vents, sunny windows). Invalid amount of sample has been ...
Specimens must be run within 10 minutes of collection to ensure accurate readings by the Analyzer. Using a specimen collected more than 10 minutes prior to insertion in the Analyzer may not generate accurate test results.
Press "RESULTS" on the Main Menu and scroll through the results. All QC tests are marked with a magnifying glass and will have the type (liquid or electronic) listed as well. Green row = pass, red row = invalid result or failure.
First, ensure that you chose the positive control first. If you did choose the positive control first, the ensure that you dispensed approximately 75 microliters of the positive control liquid into the Cartridge using the provided Specimen Transfer ...